Allergologie select, Volume 10 (2026) - 1st issue (145 - 150)

Contact allergy to a titanium port catheter: Diagnostic challenges and clinical resolution
Silas Fugger1, Alexander Kauffmann2, Heinrich Dickel1
1 Department of Dermatology, Venereology and Allergology, St. Josef Hospital, University Medical Center, Bochum, and 2 Chair for Materials Science and Engineering, Institute for Materials, Faculty of Mechanical Engineering, Ruhr University Bochum, Bochum, Germany

   

 

DOI 10.5414/ALX02637E

Abstract

Background: Titanium and its alloys are widely used in modern medical devices because of their favorable biocompatibility and mechanical properties. Allergic reactions to titanium-based implants are considered rare, and their diagnosis is hampered by the poor dermal penetration of metallic allergens. Case report: A 29-year-old female patient with seronegative myasthenia gravis requiring parenteral nutrition presented with persistent inflammatory dermatitis overlying a titanium port catheter (X-Port BARD Titan low-profile) 4 weeks after implantation in July 2023. Despite initial diagnostic uncertainty and sequential management modifications, her condition continued to deteriorate. Similar eruptions developed after re-implantation of the same titanium port model on the contralateral thorax. Rigorous patch testing incorporating standardized tape-stripping and extended reading intervals revealed positive sensitization to the titanium port body, while all other port components and an alternative plastic port system (BARD SlimPort M.R.I. Ultra Low-Profile) tested negative. Energy-dispersive X-ray spectroscopy confirmed the titanium body composition as a Ti-Al6-V4 alloy. Substitution with the plastic port catheter in February 2025 resulted in complete symptom resolution, with sustained clinical remission through July 2025. Conclusion: Although very uncommon, contact sensitization to titanium medical devices can occur and may present significant diagnostic and therapeutic challenges. Rigorous patch testing methodology, incorporating standardized tape-stripping and extended 168-hour readings, appears essential for detecting delayed-type hypersensitivity to metallic implants. Clinical resolution following biocompatible device substitution supports an allergic etiology in this case.

Author Details

Authors

Departments

  • 1 Department of Dermatology, Venereology and Allergology, St. Josef Hospital, University Medical Center, Bochum, and
  • 2 Chair for Materials Science and Engineering, Institute for Materials, Faculty of Mechanical Engineering, Ruhr University Bochum, Bochum, Germany

Address

Silas Fugger, Department of Dermatology, Venereology and Allergology, Ruhr University Bochum, Gudrunstraße 56, 44791 Bochum, Germany
Email: [email protected]

Citation

Silas Fugger, Alexander Kauffmann, and Heinrich Dickel.Contact allergy to a titanium port catheter: Diagnostic challenges and clinical resolution. Allergologie select. 2026; 10: 145-150. doi: 10.5414/ALX02637E.

###article_not_exists_msg###